4 / 4 page

Specifications
Version 1.0
SPD series
-4 –
SMARTEC BV, Delpratsingel 26
WWW.SMARTEC.NL
4811 AP Breda, The Netherlands
Tel: +31 76 5205353
Tel: +31 76 5205353
In the worst case scenario the minimum value of the span is 100 mV,
which means that the sensor has a span of 100 mV, whichcould drop
to 63mV (100 minus 27%) over the full temperature range of 100
°C.
Linearity
The linearity is defined as the maximum deviation from the best fitted
straight line.
Outline drawings and dimensions.
Other encapsulations on request.
Ordering information
Part no.
Pressure range
SPD005G
0 – 0.35 Bar (Gauge)
SPD015G
0 – 1.0 Bar
-
SPD030G
0 – 2.0 Bar
-
SPD100G
0 – 6.5 Bar
-
SPD015A
0 – 1.0 Bar (Absolute)
Connection diagram
(Top View)
pressure port (
∅
∅ 0.8)
1
2
3
12
14
vent hole
∅
∅ 0.8
6
5
4
2.54(2x)
R=1
2.8
6.8
0.5C-
∅
∅ 3
8
4
All sizes in mm.
Wetted materials: glass, ryton, silicon, fluorosilicon
and gold
4 out
+
5 exc
1 out
-
2 exc
0.5
2.54
3.5
Gauge type sensor
Absolute type sensor
0.25
0.15±0.1
11±0.5
0.8±0.2
Atmospheric
Pressure inlet
@bottom
7
7
1
4