22 / 29 page ![]() 22 PROBLEM POTENTIAL CAUSES SOLUTIONS IntelliTune fails because of a “load short” error VLIM may be too high. Reduce the current limit and rerun IntelliTune to automatically reduce VLIM. Then reset the current limit and rerun the scan. Use remote command TEC:VLIM 9 to set the VLIM to 9 V and then rerun IntelliTune. Put a resistor in series with the thermoelectric to increase the load compliance voltage. Load Fault Error is triggered A Load Fault Error may be falsely triggered when operating near maximum current and near maximum voltage. Use remote command TEC:VLIM 27 to limit the internal voltage supply to 27 V instead of the 28 V maximum. |
|
リンク URL |
| ALLDATASHEETはお客様のビジネスに役立ちますか? [ DONATE ] |
Alldatasheetは | 広告 | お問い合わせ | プライバシーポリシー | データシートへのリンク | リンク交換 | メーカーリスト All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|